Product description
- ±500Pa
- Accuracy: ± 3 Pa
- Differential pressure, volume flow
- EtherCAT P
- IP54
Networkable pressure transmitters PREMASGARD® 612x - EtherCAT P (series) PREMASGARD® 612x - ECATP with M8 plug connector (EtherCATP-coded), Bluetooth-capable, in impact-resistant plastic housing with quick-release screws, optionally with/without display and bar graph, connection piece for pressure hose (Ø 6 mm), incl. connection set ASD-06 (2 m connection hose, two pressure connection nipples, screws).
The surface-mounted sensor is used to measure positive, negative or differential pressures in clean air and gaseous media as well as for volume flow calculation via K-factor with function selection. The piezoresistive measuring element is temperature-compensated and guarantees high reliability and accuracy. The sensor is calibrated at the factory.
It is used in clean room, medical and filter technology, in ventilation and air conditioning ducts, in spray booths and commercial kitchens, for filter monitoring and level measurement or for controlling frequency converters. In direct sunlight, the weather and sun protection WS 03 (accessory) should be used.
EtherCAT P-capable pressure transmitter for industrial requirements with maximum reliability: simple integration into the PLC level via the ESI configuration file of the device, diagnostic options (such as communication error counter), extended setting options, retrievable historical data (min/max) and maintenance interval determination of the sensor. Optionally with large illuminated display (three lines, individually programmable in the 7-segment and dot-matrix area) and bargraph (seven digits, LEDs freely configurable) for graphical representation, e.g. as a traffic light display.